EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA

EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA
제품 번호: EZ3505.9700140T
Unit Price Hach에 문의

Advanced Standard Addition Technology

Ensure reliable data accuracy with advanced discontinuous and standard addition modes. This feature automatically compensates for interferences, delivering precise results even in complex or changing water conditions. Avoid compliance risks caused by inconsistent or unreliable measurements.

Designed for Low-Maintenance Operation

Frequent manual calibration and sensor cleaning increase maintenance effort and downtime. The EZ3505sc employs discontinuous and automated standard addition modes to compensate for matrix effects, stabilise potentiometric response, reduce reagent consumption, and maximise analyser uptime for reliable chloride monitoring.

Seamless System Integration

Isolated data silos hinder effective plant management. Featuring standard analogue and digital communication protocols, this analyser integrates easily into existing plant networks. Real-time visibility ensures you can make informed decisions quickly based on current process conditions.

Cost-Effective Scalability

Monitoring multiple sample points traditionally requires significant capital investment. This system supports multi-stream functionality, allowing a single unit to analyse up to 8 sample streams. You gain comprehensive process insight while optimising infrastructure costs and reducing equipment footprint.

상자에 무엇이 있습니까?

EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 2 x empty 5L Reagent Containers with Fittings (for Buffer and Standard Cl Solution)

사양

Automatic cleaning: Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Demineralised water: Internal dilution option: Rinse water and Dilution water
간섭: Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar.
검출 하한(LOD): No dilution
2 - 100 mg/L Cl-: 2 mg/L
5 - 250 mg/L Cl-: 5 mg/L
10 - 500 mg/L Cl-: 10 mg/L
20 - 1000 mg/L Cl-: 20 mg/L

Internal dilution
100 - 5000 mg/L Cl-: 100 mg/L
200 - 10000 mg/L Cl-: 200 mg/L
계측용 공기: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
교정: Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
디지털 출력: Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)

Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP

RS485 communication: Profibus DP or Modbus RTU
무게: Max. 35 kg (77 lb)
배수: Atmospheric pressure, vented, min. Ø 32 mm
범위: No dilution
2 - 100 mg/L Cl-
5 - 250 mg/L Cl-
10 - 500 mg/L Cl-
20 - 1000 mg/L Cl-

Internal dilution
100 - 5000 mg/L Cl-
200 - 10000 mg/L Cl-
보증: 12 개월
보호 등급: IP44
사이클 타임: No dilution
Default: 5 minutes
Continuous: 4 minutes

Internal dilution
Default: 10 minutes
Continuous: 7 minutes
상자에 무엇이 있습니까?: EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 2 x empty 5L Reagent Containers with Fittings (for Buffer and Standard Cl Solution)
시료 스트림 수: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
시료 압력: By external overflow vessel (open to atmospheric pressure)
시료 온도: 10 - 30 °C (50 - 86 °F)
시료 유속: 100 - 300 mL/min
시료 품질: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
시약 요구 사항: Keep between 10 - 30 °C
아날로그 출력: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
유효성 검사: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
인증: CE, ETL certified to UL and CSA safety standards, UKCA
재료: Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
전력: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
접지: Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
정밀도: Better than 2% full scale range for standard test solutions
주위 온도: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
측정 방법: Discontinuous, direct measurement by combined Ion-Selective Electrode with standard addition
  Ion-Selective Electrode (standard addition)
크기(H x W x D): 688 mm x 460 mm x 340 mm
항목: Chloride