EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA

EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA
제품 번호: EZ3005.9700140T
Unit Price Hach에 문의

Reliable, Stable Measurements

Avoid downtime from unreliable ISE data caused by temperature fluctuations. The EZ3005sc stabilises both sample and electrode using a temperature controlled flow cell, eliminating measurement errors and delivering precise, repeatable chloride results operators can trust.

Lower Maintenance Effort

Minimise downtime and service costs with the EZ3005sc. Automated 2 point calibration, reagent alerts, a rugged ISE electrode, and minimal moving parts extend maintenance intervals; so operators spend less time servicing instruments and more time on plant operations.

Reduced Operating Cost

The EZ3005sc helps optimise operating costs by reducing chemical consumption and service visits. Multi stream capability monitors up to eight sample lines, lowering cost per sampling point and extending equipment life.

Designed for easy integration into existing control systems

With a wide range of analogue and digital outputs, this analyser aligns with plant specific requirements, ensuring compatibility with existing infrastructure and immediate access to real time measurement data.

상자에 무엇이 있습니까?

EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)

사양

Automatic cleaning: Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Demineralised water: Does not apply
간섭: Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar.
검출 하한(LOD): 2 - 100 mg/L Cl-: 2 mg/L
5 - 250 mg/L Cl-: 5 mg/L
10 - 500 mg/L Cl-: 10 mg/L
20 - 1000 mg/L Cl-: : 20 mg/L
계측용 공기: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
교정: Automatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
디지털 출력: Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)

Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP

RS485 communication: Profibus DP or Modbus RTU
무게: Max. 35 kg (77 lb)
배수: Atmospheric pressure, vented, min. Ø 32 mm
범위: 2 - 100 mg/L Cl-
5 - 250 mg/L Cl-
10 - 500 mg/L Cl-
20 - 1000 mg/L Cl-
보증: 12 개월
보호 등급: IP44
사이클 타임: Default: 5 minutes
Continuous: 5 minutes
상자에 무엇이 있습니까?: EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)
시료 스트림 수: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
시료 압력: By external overflow vessel (open to atmospheric pressure)
시료 온도: 10 - 30 °C (50 - 86 °F)
시료 유속: 100 - 300 mL/min
시료 품질: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
시약 요구 사항: Keep between 10 - 30 °C (50 - 86 °F)
아날로그 출력: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
유효성 검사: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
인증: CE, ETL certified to UL and CSA safety standards, UKCA
재료: Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
전력: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
접지: Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
정밀도: Better than 2% full scale range for standard test solutions
주위 온도: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
측정 방법: Discontinuous, direct measurement by combined Ion-Selective Electrode
  Ion-Selective Electrode
크기(H x W x D): 688 mm x 460 mm x 340 mm
항목: Chloride