EZ3005 Chloride Analyser

EZ3005 Chloride Analyser
제품 번호: EZ3005.XXXXXXXX
Unit Price Hach에 문의

상자에 무엇이 있습니까?

EZ3005 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)

사양

Automatic cleaning:
Demineralised water: For rinsing
간섭: Bromide, sulphide, iodide, cyanide ions may interfere. Mercury must be absent. Ammonia and thiosulphate may interfere. Fats, oil, proteins, surfactants and tar.
검출 하한(LOD): 25 mg/L
경보: 1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
계측용 공기: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
교정: Automatic, 2-point; frequency freely programmable
디지털 출력: Optional: Modbus (TCP/IP, RS485)
모델: EZ3005
무게: 25 kg
방법 명칭: Ionselective
배수: Atmospheric pressure, vented, min. Ø 32 mm
범위: 100 - 1,000 mg/L Cl-

Optional:
25 - 250 mg/L Cl-
50 - 500 mg/L Cl-
  100 ‐ 1000 mg/L
보증: 12 개월
보호 등급: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
사이클 타임: 5 minutes
상자에 무엇이 있습니까?: EZ3005 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)
시료 스트림 수: 1, 2, 4 or 8
시료 압력: By external overflow vessel
시료 온도: 10 - 30 °C
시료 유속: 100 - 300 mL/min
시료 품질: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
시약 요구 사항: Keep between 10 - 30 °C
아날로그 출력: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
유효성 검사: Automatic; frequency freely programmable
인증: CE compliant / UL certified
재료: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
전력: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
전원공급장치: 100 - 240 VAC, 50/60 Hz
접지: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
정밀도: Better than 2% full scale range for standard test solutions
주위 온도: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
측정 방법: Discontinuous, direct measurement by combined ion-selective electrode, conform with standard methods EPA 9212 and ASTM D512-12
크기(H x W x D): 690 mm x 465 mm x 330 mm
항목: Chloride