Demineralised water:
For rinsing / dilution
간섭:
Metal ions like Lead > 10 mg/L, Zinc > 2 mg/L, Nickel > 2 mg/L, Iron > 5 mg/L, Copper > 5 mg/L. Strong oxidising agents, Cyanide, Nitrite, Phosphate (polyphosphate more than orthophosphate), Chromium, Zinc in concentrations exceeding 10 times that of Iron. Bismuth, Cadmium, Mercury, Molybdate, and Silver precipitate Phenanthroline. Polyphosphate must be absent. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
검출 하한(LOD):
≤ 2 µg/L
경보:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
계측용 공기:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
교정:
Automatic, 2-point; frequency freely programmable
디지털 출력:
Optional: Modbus (TCP/IP, RS485)
모델:
EZ2306
무게:
25 kg
배수:
Atmospheric pressure, vented, min. Ø 32 mm
범위:
0.01 - 1 mg/L Fe
0.01 - 1 mg/L Fe
Optional:
0.005 - 0.25 mg/L
0.005 - 0.5 mg/L
0.08 - 4 mg/L (with internal dilution)
0.16 - 10 mg/L (with internal dilution)
보증:
12 개월
보호 등급:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
사이클 타임:
20 min Total Fe (dilution + 5 min)
30 min Total Fe & Fe (II); Fe (II+III)
40 min Total Fe & Fe (II+III) & Fe (II) & Fe (III)
상자에 무엇이 있습니까?:
EZ2306 Iron Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 2.5L Reagent Containers with Fittings (for Reductor & Buffer Solution) and 1 x empty Glass Reagent Container with Fittings (for Colour Solution)
시료 스트림 수:
1, 2, 4 or 8
시료 압력:
By external overflow vessel
시료 온도:
10 - 30 °C
시료 유속:
100 - 300 mL/min
시료 품질:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
시약 요구 사항:
Keep between 10 - 30 °C
아날로그 출력:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
유효성 검사:
Automatic; frequency freely programmable
인증:
CE compliant / UL certified
재료:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
전력:
230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
접지:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
정밀도:
Better than 2% full scale range for standard test solutions
주위 온도:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
측정 방법:
Colorimetric measurement using TPTZ colour solution, conform with APHA 3500-Fe (B)
크기(H x W x D):
690 mm x 465 mm x 330 mm
항목:
Iron, total; Fe(II+III), total dissolved